6. Measure again once on the specimen (continued on next page!)
7. Set the desired MAX value using the key [] or []. Example: the upper specification limit 40 µm.
8. Press button [OK] to accept this value into memory.
9. OFFS appears on the display. At this stage, you can enter an offset value using the key [] or []. In the
measuring mode, this offset-value will be subtracted automatically from the readings.
Application example: having a given, constant thickness of an intermediate layer, you can set this thickness as
“offset”. In this case, only the thickness of the top coating will be displayed.
10. Press button [OK]. If you have entered an OFFS value, this will be accepted. The
mode will be left.
Finished.